Welcome to Journal of Beijing Institute of Technology
Volume 27Issue 3
.
Turn off MathJax
Article Contents
Liwei Sun, Xin Ye, Wei Fang, Zhenlei He, Yupeng Wang. Radiometric Calibration Chain Design Based on Uniform Collimated Laser Source[J]. JOURNAL OF BEIJING INSTITUTE OF TECHNOLOGY, 2018, 27(3): 385-390. doi: 10.15918/j.jbit1004-0579.17102
Citation: Liwei Sun, Xin Ye, Wei Fang, Zhenlei He, Yupeng Wang. Radiometric Calibration Chain Design Based on Uniform Collimated Laser Source[J].JOURNAL OF BEIJING INSTITUTE OF TECHNOLOGY, 2018, 27(3): 385-390.doi:10.15918/j.jbit1004-0579.17102

Radiometric Calibration Chain Design Based on Uniform Collimated Laser Source

doi:10.15918/j.jbit1004-0579.17102
  • Received Date:2017-11-08
  • A uniform collimated laser source, composed of the integrating sphere source and the off-axis parabolic, is designed. The calibration process is analyzed theoretically, and Monte Carlo ray tracing method is used to simulate and analyze the design model. The results show that when the calibration source is composed of an 8cm diameter integral sphere,10mm diameter exit port, and 1.000.mm focal length, and an off-axis angle 60° parabolic mirror, the maximum divergence angle is 7.5mrad and the uniformity of irradiance is 99.31%. Finally, the uncertainty of radiometric calibration process is analyzed, and the calibration accuracy of 0.335% for imaging spectrometer can be achieved by using the radiation calibration laser source.
  • loading
  • [1]
    Fang Wei, Yu Bingxi, Wang Yupeng, et al. Solar irradiance absolute radiometers and solar irradiance measurement on spacecraft[J]. Chinese Journal of Optics and Applied Optics, 2009, 2(1):23-28.(in Chinese)
    [2]
    Thuillier G, Foujols T, Gillotay D, et al. Solar/Solspec:scientific objectives, instrument performance and its absolute calibration using a blackbody as primary standard source[J]. Solar Physics, 2009, 257:185-213.
    [3]
    Zhao Minjie, Si Fuqi, Lu Yihuai, et al. Experimental measurement of aluminium diffuser applied to calibration system of space-borne differential optical absorption spectrometer[J]. Acta Physica Sinica, 2013, 62(24):249301.
    [4]
    Oudrari Hassan, Mcintire Jeff, Xiong Xiaoxiong, et al. JPSS-1 VⅡRS radiometric characterization and calibration based on pre-launch testing[J]. Journal of Remote Sensing, 2016, 8(41):rs8010041.
    [5]
    Uprety Sirish, Cao Changyong, Blonski Slawomir. Retrospective analysis of Suomi NPP VⅡRS radiometric bias for reflective solar bands due to operational calibration changes[J]. International Journal of Remote Sensing,2016,37(22):5472-5489.
    [6]
    Eppeldauer G P, Brown S W, Larason T C,et al. Realization of a spectral radiance responsivity scale with a laser-based source and SI radiance meters[J]. Metrologia, 2000,37:531-534.
    [7]
    Zhao Weining, Fang Wei, Sun Liwei, et al. Calibration chain design based on-ntegrating sphere transfer radiometer for SI-traceable on-orbit spectral radiometric calibration and its uncertainty analysis[J]. Chinese Physics B, 2016, 25(9):090701.
    [8]
    Pulli T, Donsberg T, Poikonen T, et al. Advantages of white LED lamps and new detector technology in photometry[J]. Light:Science & Applications, 2016,4(9):1-7.
    [9]
    Zhang Lianlian, Huang Qingmei, Li Xinyi, et al. New LED light source system with uniform illumination for line-scan CCD imaging systems[J]. Journal of Beijing Institute of Technology, 2016, 25(4):571-574.
    [10]
    Wang Yupeng, Hu Xiuqing, Wang Hongrui,et al.Standard transfer chain for radiometric calibration of optical sensing instruments with traceability in solar reflective bands[J]. Optics Precision Engineering, 2015, 23(7):1807-1812.(in Chinese)
    [11]
    Yoon H W, Gibson C E,Barnes P Y. The realization of the NIST detector-based spectral irradiance scale[J]. Metrologia, 2003, 40(1):S172-S176.
    [12]
    Sun Junqiang, Wang Menghua. On-orbit calibration of visible infrared imaging radiometer suite reflective solar bands and its challenges using a solar diffuser[J]. Applied Optics, 2015, 54(24):7210.
    [13]
    Xiong X, Erives H, Xiong S, et al. Performance of terra MODIS solar diffuser and solar diffuser stability monitor[J]. Proceedings of SPIE, 2005,5882:58820S.
    [14]
    Chen Xiangzi, Fang Wei, Wang Yupeng, et al. Main aperture area measurement of the SIAR[J]. Acta Optica Sinica, 2015,35(9):0912003.
    [15]
    Fedchak J, Carter A, Datla R. Measurement of small apertures[J]. Metrologia, 2006,43(2):S41.
    [16]
    JCGM.Evaluation of measurement data-guide to the expression of uncertainty in measurement[M].[S.l.]:Joint Committee for Guides in Metrology (JCGM), 2008.
  • 加载中

Catalog

    通讯作者:陈斌, bchen63@163.com
    • 1.

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Article Metrics

    Article views (611) PDF downloads(322) Cited by()
    Proportional views
    Related

    /

      Return
      Return
        Baidu
        map